Operator’s Manual Model 287aIonizer Performance AnalyzerP/N 0340172287a-1/100 V1.13040811 100 Housel Ave PO Box 535 Lyndonville NY
Section 5GLOSSARY Blank voltage (Main Menu Startup screen) or Polarity indicator (positive or negative voltage) or Scroll up or down listA Ref
Section 6REFERENCESESD Association (Draft) Standard Practice 3.3 for the Protection of Electrostatic Discharge Susceptible Items – Periodic Verificati
WARRANTYMonroe Electronics, Inc., warrants to the Owners, this instrument to be free from defects in material and workmanship for a period of two year
CONTENTSSpecifications ...Section 1...Page 1Specifications ...
Section 1SPECIFICATIONS:Charger: 1100 volts, selectable polarityFieldmeter:Accuracy5% of reading, 2% typical Zero Drift <4V in 90s, 2V typical
Section 2GENERAL INFORMATIONThe IPA 287a performs manual or automatic decay and balance tests for periodic verification of ionization equipment. It t
Section 3OPERATIONGetting Started, a TutorialConfidence Check – Three steps to get going —If you received this instrument directly from the factory, a
c) Press and release ‘+ CHARGE’. This will start a manual test where the test voltage is positive.(i) The instrument performs an autozero.(ii) The fl
iii) The test results are displayed at the end of the test. The results are displayed in three windows which may be scrolled with either the or bu
To display the data, press and hold the ‘SETUP’ button (same button as ‘ON-OFF’) until the display goes past ‘Off’ and shows the word ‘Set-up’, then r
Section 4MAINTENANCEPrecautions —User maintenance should be limited to keeping the instrument clean and free from physical damage. Store the instrume