Operator’s Manual Charged Plate Analyzer Model 268A-1 100 Housel Ave ⏐ PO Box 535 Lyndonville NY 14098
APPENDIX I MODEL 268A-1T CHARGED PLATE ANALYZER WITH VARIABLE TIMER OPTION VARIABLE TIMER OPTION The Model 268A-1T Charged Plate Monitor with varia
APPENDIX II Application Note APNE-0017 Plate Self Decay of Charged-Plate Monitors Introduction Charged-Plate Monitors (CPM’s) evaluate the ability
The ESD Association Standard Test Method STM3.1 recognizes that a small amount of self decay may be unavoidable. This test method allows a maximum s
WARRANTY Monroe Electronics, Inc., warrants to the Owners, this instrument to be free from defects in material and workmanship for a period of two y
TABLE OF CONTENTS Warranty Section 1 ... Specifications Section 2 ...General Information Section 3 ...
Section 1 SPECIFICATIONS CHARGED PLATE/PROBE ASSEMBLY: Designed to EOS/ESD Association standard Standard tripod mount Dimensions: 6” X 6” C
Section 2 GENERAL INFORMATION The Charged Plate Analyzer Model 268A separates into two components connected by a ten-foot cable. The top part wit
IMPORTANT NOTE In order to maintain proper operation of the floating plate, it is essential that the support insulators be kept scrupulously clean.
Section 3 OPERATION BASIC OPERATION - DISCHARGE MODE: 1. Separate sections and position as necessary. Connect to the proper power source and pr
BASIC OPERATION - PEAK MODE: Peak mode is useful for checking balance in pulse ionizing systems. The timer is not associated with these measuremen
Section 4 MAINTENANCE GENERAL: Field maintenance of Model 268A should be limited to cleaning of the insulating standoffs, cleaning of the case and