If, however there exist voltage sources in the vicinity of the sensitive electrode which are independent of the
unknown to be measured, the offsets produced will detrimentally affect the spacing independence. Such voltage
sources include contact potential differences among the internal probes parts, small specks of charged dust
particles, etc.
The null balance circuitry is provided to neutralize such offsets by applying a voltage directly to the sensitive
electrode.
C. Zero: (ZERO 2)
The instrument's zero control is a voltage source connected in the metering circuitry. It is used to overcome offsets
produced by voltage sources external to the probe.
D. Gain:
The instrument functions as a closed loop unity gain voltage follower. Its open loop gain from probe to integrator is
determined by the gain of various amplifiers. The exact gain required for optimum transient responses during a
specific measurement is determined by the probe-to-surface spacing used. In order to accommodate
measurements at various probe-to-surface separations, a front panel gain control is provided which controls the
gain of the signal amplifier.